Determinação de aminoácidos totais em amostras de plantas empregando multicomutação

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Universidade Federal de São Carlos

DOI

Abstract

In this work, development of a multicommutation system for the determination of total amino acids in samples of plants were proposed. The system is based on the complexition of amino functional groups of amino acids by ninhydrin. As system detection was used an spectrophotometer working in the visible range. The multicommutation developed system employed five solenoid valves for solution management and a peristaltic pump for the fluid propulsion. The time and sequence of switching of valves, the velocity of peristaltic pump and the data acquisition were performed through the development program in the Labview plataform using a computer with an eletronic interface. The system shows a linear reponse up to 2.0 x10-3 mol L-1, with relative standard deviation of 1.09% (n = 10), , sample troughput of 15 determinations per hour and a detection limit of 3.6 x10-5 mol L-1. The system was applied to plant samples and the results obtained using the standard addition test presented receiver between 96 and 101%.

Description

Citation

OLIVEIRA, Daniel Batista de. Determinação de aminoácidos totais em amostras de plantas empregando multicomutação. 2005. 68 f. Dissertação (Mestrado em Ciências Exatas e da Terra) - Universidade Federal de São Carlos, São Carlos, 2005.

Endorsement

Review

Supplemented By

Referenced By

Rights and licensing

Acesso Aberto